GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 74
Retrieve both factory preset values from the display
The initial factory shipment value of the Diffuse Reflection
coefficient is transferred from the display. The reflected luminance
value is then calculated using the Illuminance value measured by
the ambient light sensor of the display. The formula is:
Reflected luminance = Illuminance x Diffuse Reflection coefficient
Example: 0.04 cd/m
2
= 20 lux x 0.002 cd/m
2
per lux.
When using MD215MG, MD211G5, MD212G3 or MD302C6, this
function is disabled.
Measure and calculate
The illumination is measured automatically and the reflected
luminance value will be calculated.
When using MD215MG, MD211G5, MD212G3 or MD302C6, this
function is enabled.
IMPORTANT NOTES:
It is very important to keep the ambient light conditions during this adjustment
process and during any calibration process on the same level.
Be very careful and do not change ambient light conditions drastically around the
display during a calibration process. Otherwise an incorrect diffuse reflection
coefficient will be stored in the display and the Ambient Light Compensation
function will not work correctly.
The Diffuse Reflection coefficient typically ranges from 0.002 to 0.006 cd/m2 per lux
and highly depends on the type of anti-reflection treatment applied on the screen
surface. Changes may happen over time due to aging and/or poor surface treatment.
For a newly installed display you may use the function Retrieve both factory preset
values from the display to adjust the Ambient Light Compensation in the display
correctly. If this action reports unreasonably high Reflected Luminance values, use the
Manually adjust the Reflected Luminance value” function in addition.
Please use the function “Manually adjust the Reflected Luminance value with
utmost care. For a realistic Reflected luminance value in the actual installation
environment, a calibratable luminance meter (“telescopic type” or “distance