GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 118
White Luminance Test Error
Define the maximum deviation from the target value when the
White Luminance value is measured. The target luminance value
for white luminance calibration is used as the target value. The
maximum deviation can be set in a range from +/-10.0 cd/m
2
to
+/-100.0 cd/m
2
.
Send to client workstation
To display a dialog box on the selected display when the preset
limit is exceeded during measurement, check this checkbox.
Send to Server Check this box to send an alert to the GammaCompMD QA Server
when the preset limit is exceeded during measurement.
Select All/Deselect All Select or deselect all tick boxes for both Send to client
workstation and Send to Server.
Conformance (1) tab
Figure 101: Alert Setup Conformance (1) dialog box
LUM Test Error Define the maximum permissible value for the standard deviation
of the JND-Index intervals, which is back calculated from the
luminance characteristics after grayscale calibration. This can be
set in a range from 0.0001 to 0.9900.